Notes
Slide Show
Outline
1
Standing guard over the flame of life
  • Oxygen Monitoring in Acute Medicine
2
Life
  • 2nd law of thermodynamics
  • Entropy is time’s arrow
3
"CnHnOn"
  • CnHnOn
4
 
5
Monitoring Oxygen As a Gas
  • Inspired
  • Tracheal
  • Alveolar
  • Oxygen analyser
  • Calculated from FIO2
  • Calculated: Alveolar air equation
    End tidal oxygen analyser


6
Ideal Oxygen Analyser
  • Rapid response
  • Continuous
  • Robust
  • Low maintenance
  • Do no harm
7
Oxygen Analysers
  • Polarographic sensor
  • Fuel cell
  • Paramagnetic analyzer
    • Robust
    • Rapid response time
    • Low maintenance

8
Paramagnetic Oxygen Analyser
9
Oxygram
10
Alveolar Air Equation
  • PAO2 = FIO2 * (Pb- 47) - PaCO2
  • PAO2 = FIO2 * (Pb- 47) - PaCO2 /R
  • PAO2 = FIO2 * (Pb- 47) - PaCO2 /R(1-FIO2(1-R))


11
 
12
 
13
Ideal blood oxygen monitor
  • Indicate oxygen tension
  • Indicate oxygen supply
    • Haemoglobin concentration
    • Haemoglobin saturation
    • Blood flow
  • Rapid response
  • Continuous
  • Safe
    • Non-invasive
  • Robust
  • Low maintenance
  • Stable calibration
  • Affordable
14
Monitoring blood oxygen
  • Arterial tension
  • Arterial saturation
  • Blood gas analyser
  • Trans cutaneous
  • Calculated from tension
    Co-oximeter
    Pulse oximeter



15
Blood gas analyser
  • Polarographic
  • Measures tension
  • Calculates saturation (functional)
  • Problems
    • invasive
    • intermittent
16
Functional Saturation
  • OxyHb
  • (OxyHb+reducedHb)


  • 7g OxyHb     1g reducedHb     2g metHb


  • Functional Saturation = 7/(7+1)
  • = 87%
17
Transcutaneous gas analyser
  • Polarographic
  • Measures tension
    • persistent pulmonary hypertension
    • diaphragmatic hernia
  • Problems
    • Delicate
    • Heated
    • Repositioning
    • Recalibrating
18
Co oximeters
  • Measures haemoglobin types by absorption of light
  • To measure 4 types of Haemoglobin
    measure absorption at 4 different wavelengths
    • OHb
    • reducedHb
    • COHb
    • metHb
19
Fractional Saturation
  • OHb
  • (OHb+reducedHb+COHb +metHb)


  • 7g OxyHb     1g reducedHb     2g MetHb


  • Fractional Saturation = 7/(7+1+2)
  • = 70%
20
Oximeters
  • 1875 vonVierordt
  • 1972 Takuo Aoyagi - "Pulse Oximetry"
  • 1980's Nellcor
  • 1988 pulse oximetry mandatory in Australia
21
 
22
 
23
 
24
Oximeter errors
  • Signal artefact
  • Light absorption artefact
25
Signal artefact
  • Ambient light
  • Poor perfusion
  • Movement artefact
  • If R = 1 pulse oximeter shows 85%
26
Light absorption artefact
  • carboxyHb
  • metHb
27
 
28
 
29
Light absorption artefact
  • carboxyHb
    • error showing low functional saturation
    • error showing high fractional saturation
  • metHb
    • error toward 85%
30
Light absorption artefact
  • carboxyHb
    • reads low functional, high fractional
  • metHb
    • error toward 85%
  • fetal Hb
  • dyes
  • nail polish
  • bilirubin
  • skin pigment


31
Monitoring oxygen
  • Tissue
  • Venous tension
  • Venous saturation
  • Cytochrome oximeters
  • Blood gas analyser
  • Optical fiber oximeter


32
Monitoring Oxygen